• Gossen Metrawatt
  • Camille Bauer

Characteristic Values

SECULIFE ES PRO SECULIFE ES PRIME SECULIFE ES XTRA
Item number M695B M695U M695C
Product details page Details Details Details
Weight 1.6kg 2.3kg 2.3kg
Housing 8.6 x 23.1 x 20.3 cm / ABS plastic 19.8 x 31.8 x 57.2�cm / aluminum 19.8 x 31.8 x 57.2�cm / aluminum
Front plate Lexan / back-printed Lexan / back-printed Lexan / back-printed
Operating temperature 15°C - 30°C 15°C - 40°C 15°C - 40°C
Storage temperature -40°C - 60°C -20°C - 65°C -20°C - 65°C
Display Graphic LCD display, 128 x 64 pixels
Memory setting EEPROM, all parameters
Power supply ON < 150 mA, OFF < 40 uA
Voltage supply adapter 6 V DC, plus pole inside, 300 mA
RF current transformer (50 ?) 50 Ohm
RF current transformer attenuation 0.1:1/1:1, can be selected by the user
Voltage (RMS) input resolution 0.1 mV RMS
2.0 to 700.0�mV RMS
Voltage resolution 0.1 mV
1000.0 mV
0.1 mV
Measuring range for frequency 10kHz - 10000kHz
Accuracy of input voltage range 0,5 mV, <= 50 mV
1% rdg., 50 mV, up to 1 MHz
3% rdg., 50 mV, 1 to 10 MHz
Input, resistor 50
Current (with 0.1:1 CT) resolution 1 mA
700.0 mA RMS
Current (with 1:1 CT) resolution 0.1 mA
70.00 mA RMS
Power resolution 0,1
999,9
Crest factor resolution 0.1 0.1 0.1
Maximum input voltage 33V
Screen size 14.5 cm QVGA 18-bit color touchscreen
Input connections 1 Ethernet
1 PS/2keyboard
1 serial
3 USB
4 mm I/O safety sockets
Mouse
Output connections 1 BNC oscilloscope
25-pin Hypertronics footswitch connection
A/D resolution 14
A/D speed 64
Bandwidth 50kHz - 10000kHz
Accuracy of measurement 1% rdg.
Current range lower range 2mA - 700mA
Current range upper range 20mA - 7000mA
Current resolution (lower - upper range) RMS 0.1mA - 1mA
Power range (watts) 500W
Current resolution (watts) 0.1W
Crest factor range 1.4 - 500
Input voltage range 0.20 to 70.0 mV RMS (lower range)
20 to 700.0 mV RMS (upper range)
mV peak / peak-peak range 0 - 1
Internal relays for settings and selection of loads 10
Range of internal loads 0Ohm - 6400Ohm
Accuracy of internal loads 1% non-inductive
Line values ??of internal loads 1 Ohm : 25 W
2 ? : 50 W
4 ? : 100 W
Remaining loads: 225 W
Load bank operating cycle 10 seconds on, 30 seconds off
Load cooling

Dialogue, Information and Actions: