• Gossen Metrawatt
  • Camille Bauer

Characteristic Values

SECUTEST ST PRO SECUTEST S2 N+10 SECULIFE ST SECUTEST ST PRO BT COMFORT SECUTEST ST BASE SECUTEST SIII+ H SECUTEST SIII+
Item number M707B M7010-V032 M6930 M707C M707A M7010-V013 M7010
Product details page Details Details Details Details Details Details Details
Test sequences per VDE 0544-4 (welding equipment) x x x
Relocation management for site data (moving test objects) x x
HTML report x x x
Test sequences per IC 62353 (medical devices) x x x x x x
Data transmission to PC via BT/USB x
Supply voltage 90 - 250 115 - 230 115 - 230 90 - 250 90 - 250 115 - 230 115 - 230
Test sequences for testing PRCDs x x x
Documentation software x x x
Extension for testing and managing medical devices x x x x x x
Test sequences per IEC 60601 (medical devices)
Printer Connection USB-A USB-A USB-A
Type of interface USB-B RS-232 USB-B
Indication Graphic Graphic Graphic Graphic Graphic Graphic Graphic
Automatic test sequence x x x x x x x
Limit value indication x x x x x x x
Measured value memory x x x x x x x
Deadlines management Optional No Yes Yes Optional No No
Active testing x x x x x x x
Passive testing x x x x x x
Protective ground resistance measuring range 0Ohm - 30Ohm 0Ohm - 30Ohm 0.001Ohm - 2.1Ohm 0Ohm - 30Ohm 0.001Ohm - 30Ohm 0Ohm - 30Ohm 0.001Ohm - 2.1Ohm
Isolation resistance measuring range 0.01MOhm - 300MOhm 0.05MOhm - 300MOhm 0.001MOhm - 310MOhm 0.01MOhm - 300MOhm 1MOhm - 300MOhm 0.05MOhm - 300MOhm 0.005MOhm - 300MOhm
Protective conductor current x x x x x x x
Contact current x x x x x x x
Differential current method x x x x x x x
Direct method x x x x x x x
Alternative method x x x x x x x
Consumption current measurement x x x x x x x
Power measuring x x x x x x x
Voltage measurement x x x x x x x
Connection for external measuring adapter x x x x x x x
Testing of enlarging conduits x x x x x x x
Testing with 2 probes Yes Yes Optional Yes No Yes Yes
Interface x x x x
Printer connection x x x
Printer integrated
Suitable for use by an instructed person x x x x x x x
Connection detection x x x x x x x
Description of RFID x x
DAkkS calibration certificate x x x x x x x
Printing barcodes x x x x x
Import of data Yes No No Yes No No
Integrated rubber protection x x x
Measurement of time to trip of (P)RCDs x x
Save measurement series x x x
Test current 10A x x x x x x
remote compatible Yes Yes No Yes Optional Yes Yes
Safety class detection x x x x x x x
Management of location data Optional No No Yes Optional No No
Programmable test sequences Yes No No Yes Optional No No
Frequency of supply voltage 50 - 400 50 - 60 50 - 60 50 - 400 50 - 400 50 - 60 50 - 60
High-voltage test 500V - 4000V
CC115226 Optional
CC115227 Optional

Dialogue, Information and Actions: