• Gossen Metrawatt
  • Camille Bauer

MM007: No Results for Thermocouple Temperature Measurement

The type K temperature probe with the designation Z3431-X reads out 1 mV per °C at the output.

For measurements this means:

Connect the probe to the voltage measurement sockets and select the mV DC measuring range.
The following results, for example, are obtained after measurement: 24.5 mV which corresponds to 24.5° C.
The number is correct. The unit of measure just has to be changed mentally.
This applies to the STARLINE range as well!

Dialogue, Information and Actions: